Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual 4-33
4
SAMPLE TREND OUTPUT
07/21/08 09:56:08 SpO2=000 PR=000 PI=00.00 EXC=820:OffPat,SET
07/21/08 09:56:10 SpO2=000 PR=000 PI=00.00 EXC=828:Search,OffPat,SET
07/21/08 09:56:12 SpO2=097 PR=069 PI=04.69 EXC=800:SET
07/21/08 09:56:14 SpO2=096 PR=074 PI=02.28 EXC=C00:LowSigIQ,SET
07/21/08 09:56:16 SpO2=098 PR=078 PI=03.64 EXC=800:SET
07/21/08 09:56:18 SpO2=000 PR=000 PI=00.00 EXC=800:SET
07/21/08 09:56:20 SpO2=000 PR=000 PI=00.00 EXC=820:OffPat,SET
07/21/08 09:56:22 SpO2=096 PR=078 PI=02.68 EXC=800:SET
Backlight Operation
The backlit LCD screen of the Radical-7 Pulse CO-Oximeter Handheld can be set to five
levels of illumination when the Radical-7 Pulse CO-Oximeter operates as a standalone Pulse
CO-Oximeter. The Radical-7 Pulse CO-Oximeter temporarily indicates the illumination level on
the display following a change in illumination level. To select the level of illumination, simply press
the Backlight button located on the front panel of the Handheld.
When the Radical-7 Pulse CO-Oximeter Handheld instrument is released from the Docking
Station, the illumination of the LCD screen automatically reverts to the lowest level to conserve
battery power. To select a different level of illumination, press the Backlight button again. In the
Handheld mode, four levels of illumination are available.
When the Handheld instrument is re-attached to the Docking Station, as well as when the
Radical-7 Pulse CO-Oximeter is powered on in the Standalone configuration, the backlight is
automatically set to the maximum illumination when the instrument is AC line powered.
Operation