7-2 Radical-7 Color Display Signal Extraction Pulse Oximeter Operator’s Manual Radical-7 Color Display Signal Extraction Pulse Oximeter Operator’s Manual 7-37-3
7
specifications
Display/Indicators
Data display: %SpO
2
, %SpCO, %SpMet, pulse rate, pleth waveform, alarm status, trends, status mes-
sages, Signal IQ, perfusion index, APOD and FastSat
Display update rate: 1 second
Display Color: Multi-color
Ty p e : Backlit LCD
Pixels: 480 x 272 dots
Dot Pitch: 0.25 mm
Output Interface
SatShare (RDS-1, RDS-1B)
Serial RS-232 (RDS-1, RDS-1B, RDS-3)
Nurse Call/Analog Output (RDS-1, RDS-1B, RDS-3)
Philips Vuelink, Spacelabs Universal Flexport, RadNet, RadLink (RDS-1, RDS-1B, RDS-3)
Compliance
EMC Compliance: EN60601-1-2, Class B
Equipment Classification: IEC 60601-1 / UL 60601-1
Type of Protection Class 1 (on AC power), Internally powered (on battery power)
Degree of Protection-Patient Cable: Type BF-Applied Part
Degree of Protection-SatShare Cable: Type CF-Applied Part
Mode of Operation: Continuous
1 The Radical-7 with LNOP Adt sensors has been validated for no motion accuracy in human blood studies on healthy adult male and
female volunteers with light to dark skin pigmentation in induced hypoxia studies in the range of 70-100% SpO
2
against a laboratory
Pulse CO-oximeter and ECG monitor. This variation equals plus or minus one standard deviation. Plus or minus one standard deviation
encompasses 68% of the population weight.
2
The Radical-7 with LNOP Neo sensors has been validated for no motion accuracy in human blood studies on healthy adult male and
female volunteers with light to dark skin pigmentation in induced hypoxia studies in the range of 70-100% SpO
2
against a laboratory
Pulse CO-Oximeter and ECG monitor. This variation equals plus or minus one standard deviation. Plus or minus one standard devia-
tion encompasses 68% of the population weight. 1% has been added to the saturation accuracy to account for the effects of fetal
hemoglobin. This variation equals plus or minus one standard deviation, which encompasses 68% of the population.
3
The Radical-7 with LNOP Adt sensors has been validated for motion accuracy in human blood studies on healthy adult male and
female volunteers with light to dark skin pigmentation in induced hypoxia studies while performing rubbing and tapping motions, at 2
to 4 Hz at an amplitude of 1 to 2 cm and a non-repetitive motion between 1 to 5 Hz at an amplitude of 2 to 3 cm in induced hypoxia
studies in the range of 70-100% SpO
2
against a laboratory Pulse CO-Oximeter and ECG monitor. This variation equals plus or minus
one standard deviation. Plus or minus one standard deviation encompasses 68% of the population.
4 The Radical-7 with SatShare has been validated for low perfusion accuracy in bench top testing against a Biotek Index 2 simulator and
Masimo’s simulator with signal strengths of greater than 0.02% and a % transmission of greater than 5% for saturations ranging from
70 to 100%. This variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of
the population.
5 SpCO and SpMet accuracy has been validated on healthy adult male and female volunteers with light to dark skin pigmentation in
the range of 0% - 40% SpCO and 0% - 15% SpMet against a laboratory Pulse CO-Oximeter. This variation equals plus or minus one
standard deviation, which encompasses 68% of the population. The SpCO accuracy and SpMet accuracy have not been validated
under motion conditions.
6 This represents approximate run time with the backlight at minimum and Power Save mode on, using a new, fully charged battery.
7 If the batteries are to be stored for extended periods of time, it is recommended that they be stored between -20°C to +30°C, and at a
relative humidity less than 85%. If stored for a prolonged period at environmental conditions beyond these limits, overall battery capacity
may be diminished, and lifetime of the batteries may be shortened.
8 With FastSat the averaging time is dependent on the input signal. For the 2 and 4 second settings the averaging time may range from
2-4 and 4-6 seconds, respectively.
9 Maximum sensitivity mode fixes perfusion limit to 0.02%.
10 Masimo Rainbow SET technology with LNOP, LNOPv and LNCS sensors has been validated for pulse rate accuracy for the range of
25-240 bpm in bench top testing against a Biotek Index 2 simulator. This variation equals plus or minus one standard deviation which
encompasses 68% of the population.
Docking Station (RDS-1B):
Type: NiMH
Capacity: 10 hours
6
Charging time: 6 hours
ENVIRONMENTAL
Operating Temperature: 41°F to 104°F (5°C to 40°C)
Storage Temperature: -40°F to 158°F (-40°C to +70°C)
7
Operating Humidity: 5% to 95%, non-condensing
Operating Altitude:
500 mbar to 1060 mbar pressure
-1000 ft to 18,000 ft (-304 m to 5,486 m)
PHYSICAL CHARACTERISTICS
Dimensions:
Handheld: 8.9” x 3.3” x 2.1” (22.6 cm x 8.4cm x 5.3 cm)
Standalone: 3.5” x 10.5” x 7.7” (8.9 cm x 26.7cm x 19.6cm)
Weight
Handheld: 1.3 lbs. (0.59 kg)
Docking Station (RDS-1, RDS-2, RDS-3): 2.5 lbs. (1.14 kg)
Docking Station (RDS-1B): 4.11 lbs (1.86 kg)
Standalone (RDS-1, RDS-2, RDS-3): 3.8 lbs. (1.73 kg)
Standalone (RDS-1B): 5.4 lbs. (2.45 kg)
Trending
72 hours of trending at 2 second resolution, up to 18 days of trending at 10 second resolution, output to
serial printer or other serial devices
Mode
Averaging mode: 2, 4, 8,10, 12, 14 or 16 seconds
8
Sensitivity: Normal and Maximum
9
and APOD
Alarms
Audible and visual alarms for high low saturation and pulse rate (SpO
2
range 1-100%, pulse rate range
25-240 bpm, SpCO 1%-99%, SpMet 1%-99.9%)
Sensor condition, system failure and low battery alarms
High Priority:
571 Hz tone, 5 pulse burst, pulse spacing: 0.250s,
0.250s, 0.500s, 0.250s, repeat time:10s
Medium Priority:
550 Hz tone, 3 pulse burst, pulse spacing: 0.375s,
0.375s, repeat time: 7s
Low Priority: 500 Hz tone, 1 pulse burst, repeat time: 5s
Alarm Muted reminder:
500Hz tone, 2 pulse burst, pulse spacing 0.375s,
repeat time: 3min.
Alarm Volume:
High Priority: 70 dB (min),
Medium Priority: 70 dB (min),
Low Priority: 45 dB (min)
specifications
7-2
7