8-2 Radical-7 Color Display Signal Extraction Pulse Oximeter Operator’s Manual Radical-7 Color Display Signal Extraction Pulse Oximeter Operator’s Manual 8-38-3
8
sensors & patient cablessensors & patient cables
8-2
8
Masimo SpO
2
Sensors
The Radical-7 Pulse CO-Oximeter may use standard Masimo LNOP, LNOPv and LNCS
SpO
2
sensors, when used with Red PC, Red LNC or Rainbow Patient Cables respec-
tively.
Select the appropriate patient cable to attach the LNOP or LNCS sensor to the device.
RED REUSABLE SENSORS
(Red DC-I and DC-IP sensors must be used in conjunction with Red PC cables)
Masimo Red sensors can be used with the Radical-7 Pulse CO-Oximeter to enable measurement of
SpO
2
only. Red sensors will only function with oximeter devices equipped with Masimo Rainbow SET
technology.
SENSOR
Weight
Range
Saturation Accuracy Pulse Rate Accuracy Low Perfusion Accuracy
No Motion Motion No Motion Motion Saturation Pulse Rate
DCI > 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
DCIP 10 - 50 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
LNOP
®
REUSABLE SENSORS
(LNOP sensors must be used in conjunction with Red PC cables)
SENSOR
Weight
Range
Saturation Accuracy Pulse Rate Accuracy Low Perfusion Accuracy
No Motion Motion No Motion Motion Saturation Pulse Rate
LNOP DCI > 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
LNOP DCIP 10 - 50 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
LNOP YI > 1 kg ± 2% ± 3% ± 3 bpm ± 5 bpm N/A N/A
LNOP TC-I > 30 kg ± 3.5% N/A ± 3 bpm N/A ± 3.5% ± 3 bpm
LNOP DC-195 > 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
LNOP TF-I > 30 kg ± 2% N/A ± 3 bpm N/A ± 2% ± 3 bpm
NOTE: The LNOP TF-I and TC-I sensors were not validated under motion conditions.
Masimo Rainbow Sensors
Masimo Rainbow sensors must be used with the Radical-7 Pulse CO-Oximeter to enable
measurement of Carboxyhemoglobin (SpCO) and Methemoglobin (SpMet). Rainbow sen-
sors will only function with instruments containing Masimo Rainbow SET Technology or
licensed to use Rainbow compatible sensors.
Rainbow sensors connect to the device directly or with a patient cable.
RAINBOW REUSABLE SENSORS
SpO
2
and pulse rate accuracy for the Rainbow sensors is specified in the following table.
SENSOR
Weight
Range
Saturation
Accuracy
Pulse Rate
Accuracy
Low Perfusion
Accuracy
SpCO
Accuracy
SpMet
Accuracy
No
Motion
Motion
No
Motion
Motion Saturation
Pulse
Rate
No
Motion
No
Motion
DCI > 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
± 3% ± 1%
DCIP 10 - 50 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
± 3% ± 1%
RAINBOW ADHESIVE SENSORS
SpO
2
and pulse rate accuracy for the Rainbow sensors is specified in the following table.
SENSOR
Weight
Range
Saturation
Accuracy
Pulse Rate
Accuracy
Low Perfusion
Accuracy
SpCO
Accuracy
SpMet
Accuracy
No Motion Motion
No
Motion
Motion Saturation
Pulse
Rate
No Motion No Motion
R25-L
> 30 kg
60 - 80% ± 3%
70 - 100% ± 2%
± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm ± 3.5% ± 1%
R20-L
3 - 10 kg
60 - 80% ± 3%
70 - 100% ± 2%
± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm ± 3.5% ± 1%
10 - 30
kg
60 - 80% ± 3%
70 - 100% ± 2%
± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm ± 3.5% ± 1%