18 Radical Signal Extraction Pulse Oximeter Operator’s Manual
3
ALARM VOLUME Set to pre-power down setting
PULSE BEEP VOLUME Set to pre-power down setting
TREND DISPLAY PARAMETERS Set to pre-power down setting
TREND PERIOD Set to pre-power down setting
satshare setup
The Radical Pulse Oximeter has been proven to be accurate during patient motion and low
perfusion conditions. Saturation and pulse rate values from the Radical may be displayed on
a multiparameter monitor through the SatShare feature.
The SatShare feature provides an ideal, simulated waveform corresponding to the measured
saturation and pulse rate values determined by the Masimo SET technology.This waveform
may be used to display these values on multiparameter monitors through the
multiparameter monitor’s oximetry sensor or input connector.
It is recommended that the Radical is positioned close to the multiparameter monitor with
the Radical screen visibly displaying the plethysmographic waveform and the saturation and
pulse rate measurements.
SatShare Setup:
1. Select the SatShare cable that is appropriate for the multiparameter monitor that is
being connected. Check the Masimo web site at www.masimo.com for the latest list
of available SatShare cables and validated instruments.
2. Connect the labeled end of the cable to the SatShare Cable Connector port on the
back of the Docking Station. Tighten the connector screws for a secure connection.
3. Connect the other end of the SatShare cable either to the sensor connector of the
multiparameter monitor’s SpO
2
cable or directly to the SpO
2
connector on the monitor.
set-up
Comentários a estes Manuais