Radical Signal Extraction Pulse Oximeter Operator’s Manual 53
7
specifications
ANALOG OUTPUT
Radical can interface with various analog recording devices and/or strip chart recorders
through its Analog Output connector located on the back of the Docking Station. SpO
2
and
pulse rate parameters are output simultaneously and continuously. The
output signals vary from approximately 0 to 1 volt in a linear fashion.
NOTE: The actual SpO
2
and Pulse Rate voltages will vary from approximately 40mV to
980mV in 3.9mV steps.
SpO
2
0 volts represents either 0% SpO
2
or 50% SpO
2
. In either case, 1 volt is used to represent
100% SpO
2
. The range mapping for %SpO
2
can be selected by accessing the menu
system under Output-Analog Output Mode.
PULSE RATE
For Pulse Rate, 0 volts represents 0 BPM, while 1volt is used to represent 250 BPM.
CALIBRATION
For measurement device calibration purposes, the analog output signals can be set to either
0 volts or 1 volt in the menu system under Output/Analog Output Mode.These settings affect
both channels. Calibrate your analog recording system to those levels before use.
NURSE CALL
The nurse call feature is available when Radical is operating in its standalone configuration.
The nurse call feature on the Radical Pulse Oximeter is based on the relay closing or
opening depending on alarm events. For maximum flexibility, either normally open (pin 6)
or normally closed (pin 7) signals are available. Only qualified personnel should connect
one of these two signals and common (pin 12) to a hospitals nurse call system. During an
alarm condition, the normally open pin will be connected to the common pin and the
normally closed will be disconnected.
The nurse call relays have the following electrical specifications per switch:
PARAMETER SPECIFICATION
Max Voltage 100VDC or AC
peak
Max Current 100mA
WARNING: THE NURSE CALL FEATURE IS DISABLED WHEN THE AUDIBLE
ALARMS ARE SILENCED.