22 Radical Signal Extraction Pulse Oximeter Operator’s Manual
4
11. Verify that the patient alarms are functional by setting the high and low SpO
2
and
pulse rate alarm limits beyond the patient readings.
■
An alarm tone sounds.
■
The violated alarm limit and reading flash on the display.
■
The red alarm indicator flashes on the Docking Station (standalone operation).
12. Verify the sensor alarms are functional by removing the sensor from the sensor site.
■
SENSOR OFF appears in the message area of the graphic display.
■
The alarm tone sounds.
■
The alarm indicator flashes.
■
Disconnect the sensor from the patient cable or oximeter.
■
Confirm that “NO SENSOR” appears in the message area of the graphic display.
13. Verify alarm silence operation.
■
Create an alarm condition by lowering the SpO
2
or pulse rate high alarm
limits beyond the patient readings.
■
Press the Alarm Silence button.
■
The alarm tone ceases for the displayed amount of time.
14. To begin patient monitoring:
■
Adjust the alarm limits.
■
Adjust the alarm volumes.
■
Adjust the pulse beep volume.
15. Verify the sensor is on correctly and that the measured data is appropriate, see
Section 4, Successful SpO
2
Monitoring.
16. Monitor the patient.
17. After monitoring is complete, remove the sensor from the patient and store or
dispose of the sensor according to governing rules. See the Directions for Use of
the sensor.
18. Press and hold the Power/Standby Button for 2 seconds to turn the oximeter off.